Table 2

Monitor function results from the 10-week study period

Monitor function checkedOutcome
N/N (%) (lower 97.5% confidence bound)
SensitivityPositive predictive value
Inhaler actuations
Total2558/2560 (99.9%) (99.7)2558/2558 (100.0%) (99.9)
 Low-use actuations1600/1600 (100.0) (99.8)
 High-use actuations958/960 (99.8) (99.2)*
Inhaler insertion and removal
 Insertion29/30 (96.7) (82.8)†29/29 (100.0) (88.1)
 Removal9/10 (90.0) (55.5)9/9 (100.0) (66.4)
  • Sensitivity is the proportion of diary actuations electronically recorded by the monitor expressed as a percentage. Positive predictive value is the proportion of monitor recorded actuations that were recorded in the diary expressed as a percentage (lower 95.7% confidence bound).

  • All data from week 6 was 1 h discrepant (to within 7 min), in keeping with the time prior to the start of the New Zealand daylight savings period.

  • *Monitor 8 and monitor 17 each missed one actuation during high use.

  • †In monitor 9, there was a failure to record one insertion and one removal.