RT Journal Article SR Electronic T1 Three-month validation of a turbuhaler electronic monitoring device: implications for asthma clinical trial use JF BMJ Open Respiratory Research JO BMJ Open Resp Res FD British Thoracic Society SP e000097 DO 10.1136/bmjresp-2015-000097 VO 2 IS 1 A1 Janine Pilcher A1 Philippa Shirtcliffe A1 Mitesh Patel A1 Steve McKinstry A1 Terrianne Cripps A1 Mark Weatherall A1 Richard Beasley YR 2015 UL http://bmjopenrespres.bmj.com/content/2/1/e000097.abstract AB Background Electronic monitoring of inhaled asthma therapy is suggested as the ‘gold standard’ for measuring patterns of medication use in clinical trials. The SmartTurbo (Adherium (NZ) Ltd, Auckland, New Zealand) is an electronic monitor for use with a turbuhaler device (AstraZeneca, UK). The aim of this study was to determine the accuracy of the SmartTurbo in recording Symbicort actuations over a 12-week period of use.Methods Twenty SmartTurbo monitors were attached to the base of 20 Symbicort turbuhalers. Bench testing in a research facility was undertaken on days 0, 5, 6, 7, 8, 9, 14, 21, 28, 56 and 84. Patterns of ‘low-use’ (2 sets of 2 actuations on the same day) and ‘high-use’ (2 sets of 8 actuations on the same day) were performed. The date and time of actuations were recorded in a paper diary and compared with data uploaded from the SmartTurbo monitors.Results 2800 actuations were performed. Monitor sensitivity was 99.9% with a lower 97.5% confidence bound of 99.6%. The positive predictive value was 99.9% with a 97.5% lower confidence bound of 99.7%. Accuracy was not affected by whether the pattern of inhaler use was low or high, or whether there was a delay in uploading the actuation data.Conclusions The SmartTurbo monitor is highly accurate in recording and retaining electronic data in this 12-week bench study. It can be recommended for use in clinical trial settings, in which quality control systems are incorporated into study protocols to ensure accurate data acquisition.