AM801 | CMS50D | MD300C19 | MD300C29 | PC-60F | Overall | |
+ve cases | 38 | 40 | 28 | 25 | 28 | 159 |
-ve cases | 148 | 152 | 157 | 150 | 149 | 756 |
Total | 186 | 192 | 185 | 175 | 177 | 915 |
FPR | 10.8% | 21.1% | 7.6% | 10.7% | 6.0% | 11.2% |
FNR | 15.8% | 17.5% | 25.0% | 28% | 42.9% | 24.5% |
PPV | 66.7% | 50.8% | 63.6% | 52.9% | 64.0% | 58.5% |
NPV | 95.7% | 94.5% | 95.4% | 95.0% | 92.1% | 94.5% |
The rows titled ‘+ve cases’ and ‘−ve cases’ show the total number of data pairs with a respective positive or negative reading from the reference device. The FPR, FNR, PPV and NPV show, respectively, the observed FPR, FNR, PPV and NPV for each device.